Detail of Recall

Type of Product:Device
Medsafe Reference:17146
Brand Name:Siemens Dimension Vista Chemistry 1 Calibrator
Model:Catalogue Number: KC110
Affected:Lot Number: 3GM081
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL lot 3GM081. Investigation by Siemens confirms that QC and patient samples show low recovery of MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/8/2014