Detail of Recall

Type of Product:Device
Medsafe Reference:17127
Brand Name:AMS AdVance XP Male Sling System - Needle passers
Model:720163-01
Affected:all lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:American Medical Systems Inc
Issue:As a result of routine periodic product shelf life testing, manufacturer has identified that the sterile barrier of some samples of packaging of the Needle Passer components supplied with the AdVance® XP Male Sling System, may have potentially been breached and therefore do not consistently meet the three (3) year shelf life stated on the product label. Further testing confirms that the packaging of the Needle Passers continues to meet the requirements for the one (1) year shelf life.
During implant procedures of the AdVance® XP Male Sling System, the two Needle Passers (left and right sides), are used by the Surgeon to assist transobturator passage while the placing and positioning the Sling. These tools are disposable and indicated for one-time-use only. There is a potential for an increased risk of patient infection due to breach in the sterile barrier (packaging) of the Needle Passer components.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/8/2014