Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17113 |
| Brand Name: | Gambro Ultrafilter U9000 |
| Model: | Catalogue Number: 112062 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Gambro Pty Ltd (NZ) 5B Target Court Glenfield Auckland 0753 |
| Contact Information: | 021 966 006 |
| Manufacturer: | Gambro Dialysatoren GmbH & Co KG |
| Issue: | The manufacturer has become aware of the potential for external fluid leakage from the Ultrafilter U9000 as a result of multiple disinfections using a sodium carbonate-based solution. Leakage could lead to excessive fluid removal in patients. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/8/2014 |