Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17112
Brand Name:Olympus OTV-S190 HDTV Visera Elite Video Processor
Model:OTV-S190
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:Olympus CLV-S190 Xenon Visera Elite Light Source
Model:CLV-S190
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:Olympus CLV-190 Xenon Exera III Light Source
Model:CLV-190
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:Olympus UHI-4 High Flow Insufflator Unit
Model:UHI-4
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:Olympus UPD-3 Endoscope Postion Detecting Unit
Model:UPD-3
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:Olympus CV-170 Video Image Processor
Model:CV-170
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:Olympus CV-190 HDTV Exera III Video Processor
Model:CV-190
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:Olympus 3DV-190 3D Visualisation Unit
Model:3DV-190
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The power switch of the above mentioned models may become stuck after extended use and the following may occur:
1) When the power switch is stuck whilst the unit is ON, the unit cannot be turned OFF using the power switch on the model.
2) When the power switch is stuck whilst the unit is OFF, the unit cannot be turned ON using the power switch on the unit.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/8/2014
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /