Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17112 |
| Brand Name: | Olympus OTV-S190 HDTV Visera Elite Video Processor |
| Model: | OTV-S190 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Olympus (NZ) Ltd 15D Paul Matthews Road Albany Auckland |
| Contact Information: | 0508 659 6787 |
| Manufacturer: | Olympus Medical Systems Corporation |
| Brand Name: | Olympus CLV-S190 Xenon Visera Elite Light Source |
| Model: | CLV-S190 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Olympus (NZ) Ltd 15D Paul Matthews Road Albany Auckland |
| Contact Information: | 0508 659 6787 |
| Manufacturer: | Olympus Medical Systems Corporation |
| Brand Name: | Olympus CLV-190 Xenon Exera III Light Source |
| Model: | CLV-190 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Olympus (NZ) Ltd 15D Paul Matthews Road Albany Auckland |
| Contact Information: | 0508 659 6787 |
| Manufacturer: | Olympus Medical Systems Corporation |
| Brand Name: | Olympus UHI-4 High Flow Insufflator Unit |
| Model: | UHI-4 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Olympus (NZ) Ltd 15D Paul Matthews Road Albany Auckland |
| Contact Information: | 0508 659 6787 |
| Manufacturer: | Olympus Medical Systems Corporation |
| Brand Name: | Olympus UPD-3 Endoscope Postion Detecting Unit |
| Model: | UPD-3 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Olympus (NZ) Ltd 15D Paul Matthews Road Albany Auckland |
| Contact Information: | 0508 659 6787 |
| Manufacturer: | Olympus Medical Systems Corporation |
| Brand Name: | Olympus CV-170 Video Image Processor |
| Model: | CV-170 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Olympus (NZ) Ltd 15D Paul Matthews Road Albany Auckland |
| Contact Information: | 0508 659 6787 |
| Manufacturer: | Olympus Medical Systems Corporation |
| Brand Name: | Olympus CV-190 HDTV Exera III Video Processor |
| Model: | CV-190 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Olympus (NZ) Ltd 15D Paul Matthews Road Albany Auckland |
| Contact Information: | 0508 659 6787 |
| Manufacturer: | Olympus Medical Systems Corporation |
| Brand Name: | Olympus 3DV-190 3D Visualisation Unit |
| Model: | 3DV-190 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Olympus (NZ) Ltd 15D Paul Matthews Road Albany Auckland |
| Contact Information: | 0508 659 6787 |
| Manufacturer: | Olympus Medical Systems Corporation |
| Issue: | The power switch of the above mentioned models may become stuck after extended use and the following may occur: 1) When the power switch is stuck whilst the unit is ON, the unit cannot be turned OFF using the power switch on the model. 2) When the power switch is stuck whilst the unit is OFF, the unit cannot be turned ON using the power switch on the unit. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/8/2014 |