Detail of Recall

Type of Product:Device
Medsafe Reference:17075
Brand Name:Mathys RM Classic Reamers
Model:Article Numbers: 3.14.251, 3.14.252, 3.14.256, 3.14.257, 3.14.258, 3.14.259, 3.40.501, 55.02.1901, 55.02.1903, 5502.00.0, 3.14.545
Affected:All lot numbers
Software version:
Recalling Organisation:Mathys New Zealand Ltd
46 Hillside Road
Wairau Valley
AUCKLAND 0627
Contact Information:0800 628 497
Manufacturer:Mathys Medical Ltd
Issue:The validated process for the flexible shaft provides for a combined manual and mechanical cleaning/disinfection procedure. If the mandatory manual process as described in the processing instruction cannot be ensured/performed, instruments by Mathys Ltd with flexible shafts may not be used.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/8/2014