Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:17069
Brand Name:Sorin 3T Heater Cooler with Remote
Model:16-02-80
Affected:
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Sorin Group Deutschland GmbH
Brand Name:Sorin S3 Heater Cooler
Model:16-02-50
Affected:
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Sorin Group Deutschland GmbH
Issue:Newly identified risk for cardiac surgery patients whereby some patients have been infected with a slow growing Mycobacterium chimaera. This organism is found in water including tap water sources.
Important to review hygiene and surgical practices in the cardiac surgery theatre. This review should include sampling and monitoring programs for water sources solution preparations and systems that use water in the cardiac surgery theatre. Among these water systems, heater cooler device(s) need strict adherence to the cleaning disinfection and maintenance according to the operator manual for Sorin devices
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/7/2014