Detail of Recall

Type of Product:Device
Medsafe Reference:17063
Brand Name:Sysmex CS-2000i, CS-2100i and CS-5100 Automated blood coagulation analysers
Model:Catalogue number - 10488064, 10488062 and 10709128 respectively
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Sysmex Corporation
Issue:The manufacturer has confirmed a carry-over risk on Sysmex CS systems due to the applications using CA-Clean I (964-0631-3) for sample probe rinse. Under certain circumstances the carry-over might impact results for Factor V with Innovin®, Factor XII with Dade® Actin® FSL, Berichrom F XIII, INNOVANCE® D-Dimer assays.
Follow-up communication
The manufacturer has issued a new protocol disc version with new and enhanced assay applications for Sysmex CS-2000i/CS-2100i Systems.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/7/2018