Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 17037 |
| Brand Name: | Baxter CoSeal Surgical Hemostatic Sealant, 2ml and 4ml |
| Model: | Product code: 934073 (2ml) & 934074 (4ml) |
| Affected: | Selected lots |
| Software version: | |
| Recalling Organisation: | Baxter Healthcare Ltd 33 Vestey Drive Mt Wellington AUCKLAND |
| Contact Information: | 0800 229 837 |
| Manufacturer: | Baxter Healthcare Corporation SA |
| Issue: | The recall is being conducted as a precautionary measure due to out of specification results observed at the 18-month time point (out of 18-month licensed shelf life to expiration) during a COSEAL stability study. The parameter that is out of specification is an indicator of possible failure for the product to gel appropriately. COSEAL's failure to gel does not represent risk for the patient's life. All other stability requirements were met. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/7/2014 |