Type of Product:Device
Medsafe Reference:17001
Brand Name:Physio Control Lifepak 1000 Defibrillator Automated
Model:LP1000
Affected:
Software version:2.42 or earlier
Recalling Organisation:Medxus Ltd
Unit 4
100 Fitgerald Avenue
Christchurch 8011
Contact Information:03 365 7403
Manufacturer:Physio-Control Corporation
Issue:Physio-Control has become aware of incidents where customers have attempted to use their LIFEPAK 1000 defibrillator and the device has shut down unexpectedly due to a very low battery. A defibrillator in this scenario has the potential to fail to deliver a shock, with the potential result that therapy is not delivered and a patient is not resuscitated.
Physio-Control has determined this to be as a result of batteries not being replaced when they have reached a low or very low state of charge as indicated in the Readiness Display on the device.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/7/2014