Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16992 |
| Brand Name: | Philips Ingenia MR Systems |
| Model: | Models: 1.5T and 3.0T |
| Affected: | Shipped between Dec-2013 and Mar-2014 |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Issue: | In the Ingenia rear cover set magnetic M10 nuts (four) were erroneously supplied to the installer instead of A4 stainless steel nuts. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/7/2014 |