Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16992
Brand Name:Philips Ingenia MR Systems
Model:Models: 1.5T and 3.0T
Affected:Shipped between Dec-2013 and Mar-2014
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:In the Ingenia rear cover set magnetic M10 nuts (four) were erroneously supplied to the installer instead of A4 stainless steel nuts.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/7/2014