| Type of Product: | Device |
| Medsafe Reference: | 16980 |
| Brand Name: | GE CARESCAPE Respiratory Modules E-sCO, E-sCOV, E-sCAiO, E-sCAiOV, E-sCAiOE, E-sCAiOVE |
| Model: | Modules E-sCO, E-sCOV, E-sCAiO, E-sCAiOV, E-sCAiOE, E-sCAiOVE |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Healthcare Finland Oy |
| Issue: | Displayed Et-/Fi O2 values may be incorrect. A random deviation from the allowed tolerance limits of an O2 sensor component in the CARESCAPE respiratory modules, the Airway Gas Option and their respective service exchange units could cause up to 50% measurement error in Et-/Fi O2 values. An incorrect Et- /Fi O2 value could lead to a potential hypoxic situation or impaired clinical decision making. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/7/2014 |