Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16973 |
| Brand Name: | Hill-Rom Affinity 4 Birthing Bed |
| Model: | Part Number: P3700B/C/D/E |
| Affected: | Manfactured between Oct-2006 and Jun-2014 |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | Hill-Rom Inc |
| Issue: | Hill-Rom would like to remind users of proper handling of the Lift-Off foot section per the Affinity Four Birthing Bed manual. Improper handling, such as allowing the foot section to be dropped repeatedly, could cause the attachment latch mechanism to become bent. If the latch mechanism is bent badly enough, it could potentially cause the installed foot section to be improperly engaged onto the bed. If the damaged foot section should disengage during use, this could result in injury to the user due to a fall. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/7/2014 |