Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16968
Brand Name:Stryker Sonopet Console
Model:Catalogue Number: 5450852000
Affected:
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Medtech K.K.
Issue:Stryker has noticed a trend of complaints regarding insufficient irrigation to the tip of the Sonopet system. Lack of irrigation can potentially lead to excessive heat at the tip.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/7/2014