Detail of Recall

Type of Product:Device
Medsafe Reference:16964
Brand Name:Abbott 5mL Reaction Vessels
Model:List Number: 04J71-20
Affected:Lots: 56944001, 57026001, 57059001, 57673001
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott Molecular Inc
Issue:The manufacturer has determined specific lot numbers of reaction vessels may have an insufficient rim. It is estimated that less than 1% of reaction vessels within these lots may be defected. These lots have been in distribution since July 2012. A reaction vessel with an insufficient rim may not be held securely in the 1mL subsystem carrier. This may cause the reaction vessel to be seated incorrectly in the carrier or drop through it.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/7/2014