Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16964 |
| Brand Name: | Abbott 5mL Reaction Vessels |
| Model: | List Number: 04J71-20 |
| Affected: | Lots: 56944001, 57026001, 57059001, 57673001 |
| Software version: | |
| Recalling Organisation: | Abbott Diagnostics Division Ground Floor, Building D 4 Pacific Rise Mt Wellington AUCKLAND 1640 |
| Contact Information: | 0800 656 233 |
| Manufacturer: | Abbott Molecular Inc |
| Issue: | The manufacturer has determined specific lot numbers of reaction vessels may have an insufficient rim. It is estimated that less than 1% of reaction vessels within these lots may be defected. These lots have been in distribution since July 2012. A reaction vessel with an insufficient rim may not be held securely in the 1mL subsystem carrier. This may cause the reaction vessel to be seated incorrectly in the carrier or drop through it. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/7/2014 |