Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16948
Brand Name:Philips MobileDiagnost wDR Mobile Digital X Ray System
Model:
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:Under specific error conditions the system executes an unintended movement:
- Operator requests movement by pressing the dead man handle switch and pushing or pulling the handle bar
- Strain gauges defect
- Operator/bystander standing close to the device
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/7/2014