Detail of Recall

Type of Product:Device
Medsafe Reference:16936
Brand Name:McKesson Horizon Cardilogy Hemodynamics system
Model:Models: PB1000 & PB2000
Affected:
Software version:Versions 12.2 HF4 and earlier
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Medcon Ltd
Issue:McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker indicating the date of the next maintenance test the need for performing annual maintenance testing is not reiterated in 12.2 HF4 and earlier versions of the Horizon Cardiology Hemo user guides. If the parameter boxes are not calibrated, the Horizon Cardiology Hemo system may not function properly or potentially affect the accuracy of clinical parameters.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/7/2014