Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16895
Brand Name:Oxoid Legionella Latex Test
Model:DR0800M
Affected:Lot: 1316151
Software version:
Recalling Organisation:Oxoid NZ
3 Atlas Place
Mairangi Bay
AUCKLAND 1333
Contact Information:09 980 6700
Manufacturer:
Brand Name:Oxoid Legionella Pneumophila Serogroup 2-14 Test
Model:DR0802M
Affected:Lot 1301909
Software version:
Recalling Organisation:Oxoid NZ
3 Atlas Place
Mairangi Bay
AUCKLAND 1333
Contact Information:09 980 6700
Manufacturer:
Issue:A technical investigation has confirmed that there is potential for a weak or negative reaction when using Oxoid Legionella pneumophila 2-14 reagent Lot:1301912 with Serogroup 9 isolates. This reagent is contained within the kits noted above. Continued use of this reagent may result in a delay to the correct characterisation of L.pneumophila Serogroup 9 isolates.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/6/2014