Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16869 |
| Brand Name: | Intelerad IntelePACS System |
| Model: | |
| Affected: | |
| Software version: | Version 3.8.1 and up |
| Recalling Organisation: | Pharmaceutical Marketing New Zealand Ltd 162 Nile Road Milford Auckland 0620 |
| Contact Information: | 09 377 0660 |
| Manufacturer: | Intelerad Medical Systems Incorporated |
| Issue: | This issue relates to the compression of images into JPEG 2000 (J2C) format. In specific circumstances, when writing image data to storage that has filled up, the image data may become truncated. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/6/2014 |