Type of Product:Device
Medsafe Reference:16856
Brand Name:Cryolife Bioglue Surgical Adhesive
Model:CRY-BG3515-5
Affected:Lot No: all lot numbers pre 14MGV024
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Artivion Inc
Issue:At the request of the FDA the manufacturer has changed the expiration dating for Bioglue Surgical Adhesive from 24 months to 18 months due to the fact that the serum albumin component monomer failed to meet internally established end of shelf-life specification. The specification limit was set based on manufacturing capability at the time the process was first established.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/6/2014