Type of Product:Device
Medsafe Reference:16855
Brand Name:Siemens AXIOM Artis X-ray system, handswitch
Model:Model: 3771750
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Brand Name:Siemens AXIOM Artis zee X-ray system, handswitch
Model:Model: 3771750
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:There is a potential problem with AXIOM Artis or Artis zee system in connection with hand switch material number 3771750 operated at the patient table to release a x-ray image. The hand switch to release an x-ray image at the patient table is equipped with a cable extension inside the table. A problem may occur if the connection between cable connection and hand switch is contaminated with an extraordinary large amount of fluid. In a worst case situation, uncontrolled release of an x-ray image may happen. If this situation occurs, the system terminates uncontrolled x-ray after a short period.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/6/2014