Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16855 |
| Brand Name: | Siemens AXIOM Artis X-ray system, handswitch |
| Model: | Model: 3771750 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Brand Name: | Siemens AXIOM Artis zee X-ray system, handswitch |
| Model: | Model: 3771750 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | There is a potential problem with AXIOM Artis or Artis zee system in connection with hand switch material number 3771750 operated at the patient table to release a x-ray image. The hand switch to release an x-ray image at the patient table is equipped with a cable extension inside the table. A problem may occur if the connection between cable connection and hand switch is contaminated with an extraordinary large amount of fluid. In a worst case situation, uncontrolled release of an x-ray image may happen. If this situation occurs, the system terminates uncontrolled x-ray after a short period. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/6/2014 |