Detail of Recall

Type of Product:Device
Medsafe Reference:16845
Brand Name:GE Revolution XR/d X-ray systems
Model:
Affected:All systems
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Brand Name:GE Proteus XR/a X-ray systems
Model:
Affected:All systems installed with collimator part numbers: 2261763, 2261765
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Hualun Medical Systems Co Ltd
Issue:If a device is improperly re-installed, the collimator may not be properly seated and may loosen and fall. Such an event may occur immediately following re-installation during a service maintenance activity, but may also occur a few months after a service activity and continued use. A fall of a collimator while the system is in use could result in an injury to a patient or operator. There was a reported incident of a collimator fall which caused a serious patient injury.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/6/2014