Detail of Recall

Type of Product:Device
Medsafe Reference:16794
Brand Name:Puritan Bennett 840 Ventilator, power supply
Model:Model 840
Affected:
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Covidien LLC
Issue:The manufacturer has received customer reports of the loss of graphical user interface (GUI) display information while the ventilator continues to provide breath support. Customer reports describe the GUI display screens (both upper and lower) becoming blank during patient use. This event interferes with the clinician's ability to view and/or change ventilator setting or clinician-set patient alarm parameters. Please note that in all cases reported to date, the ventilator continues to provide respiratory support during these events.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/6/2014