Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16779
Brand Name:Siemens Advia Centaur Systems Troponin Ultra ReadyPack
Model:Catalogue Numbers: 02789602 & 02790309
Affected:Lots numbers: 15 lots & 15 lots respectively
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics, through internal investigation of lot 084, has confirmed that the solid phase reagent in some of the ADVIA Centaur® TNI-Ultra ReadyPacks®, lot 084, is darker in appearance, creating a potential for incorrect control and patient results.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/6/2014