Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16761 |
| Brand Name: | Handicare System RoMedic ReTurn 7400 and 7500 Patient Transfer device |
| Model: | Models 7400 and 7500 |
| Affected: | Manufactured between Apr-2009 and May 2011 |
| Software version: | |
| Recalling Organisation: | Active Rehab Equipment 25 Rennie Drive Airport Oaks Auckland |
| Contact Information: | 0800 336 339 |
| Manufacturer: | Handicare AB |
| Issue: | During the period April 2009 to May 2011, Handicare AB used a wing handle in which the screw and the handle, after repeated tightening, may slide apart. This causes a small play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually this may cause mechanical fatigue in the material of the ladder. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/7/2014 |