Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16758
Brand Name:ArjoHuntleigh Tenor Patient Transfer Lift (without scale)
Model:Tenor without scale
Affected:All devices manufactured since 2004
Software version:
Recalling Organisation:Arjo New Zealand Limited
34 Vestey Drive
Mt Wellington
AUCKLAND 1060
Contact Information:0800 000 151
Manufacturer:ArjoHuntleigh AB
Issue:During use, the bolt in the connection between the spreader bar and the lifting arm jig is subjected to mechanical forces, which may lead to breakage. This is due to the design of the connection between spreader bar and lifting arm jig, which in long term use can cause metal fatigue in this part and the formation of micro-cracks in the thread of the bolt. This issue is only related to Tenor devices without scales.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Public Recall
Date Commenced:26/6/2014