Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16758 |
| Brand Name: | ArjoHuntleigh Tenor Patient Transfer Lift (without scale) |
| Model: | Tenor without scale |
| Affected: | All devices manufactured since 2004 |
| Software version: | |
| Recalling Organisation: | Arjo New Zealand Limited 34 Vestey Drive Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 000 151 |
| Manufacturer: | ArjoHuntleigh AB |
| Issue: | During use, the bolt in the connection between the spreader bar and the lifting arm jig is subjected to mechanical forces, which may lead to breakage. This is due to the design of the connection between spreader bar and lifting arm jig, which in long term use can cause metal fatigue in this part and the formation of micro-cracks in the thread of the bolt. This issue is only related to Tenor devices without scales. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Public Recall |
| Date Commenced: | 26/6/2014 |