| Type of Product: | Device |
| Medsafe Reference: | 16747 |
| Brand Name: | Cook Hemospray Endoscopic Hemostat |
| Model: | Models: HEMO-7-EU, HEMO-10-EU |
| Affected: | W3320783 (HEMO-7), and W3344464 (HEMO-10) |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | Wilson-Cook Medical Inc |
| Issue: | Cook Medical has received a small number of reports related to the inability to spray the powder for the Hemospray Endoscopic Hemostat. These reports involve difficulty spraying the Hemospray powder from the delivery system. Adverse events that may occur due to spraying difficulties include inadequate treatment. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/5/2014 |