Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16745 |
| Brand Name: | Siemens syngo RT Dosimetrist |
| Model: | |
| Affected: | |
| Software version: | Version 2.7 |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | Under special workflow conditions involving reloading data, a newly created structure is renamed incorrectly by the software as the name of the last deleted structure. If this incorrect structure is used for treatment planning, there is a potential that an incorrect target is treated. This issue is only unlikely to be detected if the incorrect structure is very similar to the correct structure. The probability of a severe mistreatment is practically impossible with good clinical practice, as considerable changes from the original treatment plan would be detected during the treatment planning and approval. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/5/2014 |