Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16745
Brand Name:Siemens syngo RT Dosimetrist
Model:
Affected:
Software version:Version 2.7
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:Under special workflow conditions involving reloading data, a newly created structure is renamed incorrectly by the software as the name of the last deleted structure. If this incorrect structure is used for treatment planning, there is a potential that an incorrect target is treated. This issue is only unlikely to be detected if the incorrect structure is very similar to the correct structure. The probability of a severe mistreatment is practically impossible with good clinical practice, as considerable changes from the original treatment plan would be detected during the treatment planning and approval.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/5/2014