Detail of Recall

Type of Product:Device
Medsafe Reference:16740
Brand Name:Philips Ingenia, Intera and Achieva MR Systems
Model:Models: Ingenia, Intera, and Achieva
Affected:
Software version:Versions: R5.1.1 and R5.1.2
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The cross reference lines of transversal images are displayed in an incorrect position on a fused image which has been generated through MobiView post processing.
The cross reference lines are correctly displayed on the unfused stations.
The problem can occur when the field of view of the clinical scan stations is below (more towards the feet) the light visor position. In such situation an error is made in calculating the position information. This causes incorrect positioning of cross-reference lines on fused sagittal images.
The hazard is only expected in spine examinations, when using a whole spine survey, while the cervical spine is skipped for the clinical scan.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/5/2014