Detail of Recall

Type of Product:Device
Medsafe Reference:16729
Brand Name:Siemens Acuson S2000 and S3000 Ultrasound System
Model:Catalogue numbers – 10041461, 10440017 and 10441730
Affected:
Software version:Version VC25
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:When using the 9L4 transducer in combination with Virtual Touch Quantification (VTq), the displayed Mechanical Index (MI) may be outside the stated tolerance of +/-15%. This occurs only between the depths of 2.5cm and 3.4cm.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/5/2014