Detail of Recall

Type of Product:Device
Medsafe Reference:16713
Brand Name:Siemens Ysio Systems
Model:10281013, 10281163
Affected:All units
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:If excessive liquid is used for cleaning, the liquid may infiltrate the generated console and cause a short circuit which may lead to an unintended release of radiation.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/5/2014