Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16696 |
| Brand Name: | Draeger Neoflow Flow Sensor and Neoflow Insert |
| Model: | Part Number: 8410179 |
| Affected: | Batch numbers preceding batch EJ-FC |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Issue: | The manufacturer has determined that the Neonatal Flow Sensor Insert can leak after connecting the plug. This was caused by a tool within the Neonatal Flow sensor insert production process. |
| Recall action type: | Recall |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/5/2014 |