Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16696
Brand Name:Draeger Neoflow Flow Sensor and Neoflow Insert
Model:Part Number: 8410179
Affected:Batch numbers preceding batch EJ-FC
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Issue:The manufacturer has determined that the Neonatal Flow Sensor Insert can leak after connecting the plug. This was caused by a tool within the Neonatal Flow sensor insert production process.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/5/2014