Detail of Recall

Type of Product:Device
Medsafe Reference:16679
Brand Name:Biomet G7 Positioning Guide Rod
Model:Part Number: 110003458
Affected:Lots: 360877, 377304, 424960, ZB130701, ZB130702, ZB130703, ZB130704, ZB130705, ZB131201, ZB140204
Software version:
Recalling Organisation:Biomet New Zealand Ltd
Contact Information:0275 522 595
Manufacturer:Biomet Inc
Brand Name:Biomet G7 Positioning Guide Post
Model:Part Number: 110003500
Affected:Lots: 360877, 469610, 881430, ZB130701, ZB130702, ZB130703, ZB130901, ZB131001, ZB131002, ZB131101, ZB140101
Software version:
Recalling Organisation:Biomet New Zealand Ltd
Contact Information:0275 522 595
Manufacturer:Biomet Inc
Issue:Biomet has initiated this action following an investigation which identified that the positioning guide rod may fracture during impaction of the acetabular shell. If the positioning guide rod fractures, then the positioning guide post may become stuck to the insertion handle.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/10/2014