Detail of Recall

Type of Product:Device
Medsafe Reference:16675
Brand Name:Philips IntelliVue Patient Monitor, lithium ion battery
Model:Battery models: M4605A and M4607A
Affected:All batteries
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Childrens Medical Ventures
Issue:On-going monitoring of quality data determined that the risk of battery failure increases with age, when a battery remains in use longer than 3 years after the date of manufacture or 500 charge-discharge cycles. Such failure may result in overheating.
Battery management and safety practices information for the batteries M4605A / M4607A are not clearly documented in the Instructions for Use for IntelliVue Patient Monitors with software releases up to and including G.0.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/5/2014