Detail of Recall

Type of Product:Device
Medsafe Reference:16672
Brand Name:Smith & Nephew Echelon Hip System
Model:All models
Affected:All lots
Software version:
Recalling Organisation:Smith & Nephew Surgical Ltd
Unit A 36 Hillside road
Wairau Valley
AUCKLAND 0760
Contact Information:0800 657 799
Manufacturer:Smith & Nephew Inc
Issue:Smith & Nephew Surgical Pty Ltd has reported seven (N=7) Medical Device Incident Reports (MDIR) of femoral stem component breakages. In all cases, the results of Smith & Nephew's investigations cited poor proximal bone support as a possible cause of implant failure.
As a result of these investigations and Smith & Nephew's health hazard evaluation of this failure mode, the ECHELON surgical technique has been revised to reflect advice provided in the device Instructions for Use (IFU).
Recall action type:Alert
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/5/2014