Type of Product:Device
Medsafe Reference:16668
Brand Name:Idoman Thermablate Endometrial Ablation System (EAS)
Model:All devices
Affected:All lots
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Idoman Teoranta
Issue:Two changes have been made to the Instructions for Use for Thermablate-EAS
1. Physician to conduct a second sounding of the uterus just after dilation and before performing hysteroscopic examination of the uterus to confirm that perforation of the uterus or the creation of a false passage has not occurred during the dilation and soundings.
2. Clarification of contraindications
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional
Date Commenced:13/5/2014