Detail of Recall

Type of Product:Device
Medsafe Reference:16654
Brand Name:Roche ProCell/CleanCell Aspiration Tube Filters
Model:Part number: 03149773001
Affected:All lots
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:The manufacturer has identified that under certain conditions the white aspiration tube/filters on the end of the ProCell and CleanCell Aspiration tubes may become loose or fall off the aspiration tube completely. This in turn may allow air to enter the aspiration tube, creating foam. The Liquid Short Sensors (LSS) are unable to detect these very small bubbles. Foam may be visible in the ProCell and CleanCell reservoirs. If foam reaches the measuring cells, this may cause discrepant results. The magnitude of the discrepancy will vary with the amount of air being aspirated, particularly for ProCell.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/5/2014