Detail of Recall

Type of Product:Device
Medsafe Reference:16641
Brand Name:Synthes SynFix-LR Implant Holder
Model:Part Number 03.802.039
Affected:All lots
Software version:
Recalling Organisation:Synthes New Zealand Ltd
507 Mt Wellington Highway
Mt Wellington
Auckland 1060
Contact Information:0800 657 734
Manufacturer:Synthes GmbH
Issue:Complaints have been received which describe the SynFix-LR implant holder breaking at the interface between the implant and holder. If an un-retrieved device fragment remains threaded into the plate, it will not be possible to properly attach the SynFix-LR Aiming Device to the implant (plate). Proper attachment of the Aiming Device to the plate is required for accurate insertion of the four SynFix-LR screws into the SynFix-LR implant (plate) and vertebral bodies. If the tip of the holder should break, the potential exists for an un-retrieved device fragment (URDF) to be left in the SynFix-LR Implant.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional
Date Commenced:6/5/2014