Brand Name:Fujifilm Synapse CV Cardio Software
Model:
Affected:
Software version:Version 4.0.8
Recalling Organisation:Fujifilm NZ Limited
2c William Pickering Drive
Albany
Auckland
Contact Information:0800 242 646
Manufacturer:Fujifilm Medical Systems USA Inc
Issue:The "Switch User" feature is an option in Windows 7 that allows another user to open a new session with his user account without requiring the previous user to log out of the existing session. It has been discovered that running the same Synapse CV Client application under more than one Windows user account simultaneously may cause outbound HL7 reports with incorrect data to be sent to the EMR.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory