Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16614
Brand Name:Medtronic Navigator Shunt Kit
Model:Model: 9733605, 9733606
Affected:17 lots & 6 lots respectively
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Brand Name:Medtronic Navigator Tumor Resection Kits
Model:Part Number: 9733607 & 9733608
Affected:13 lots and 2 lots respectively
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Brand Name:Medtronic Navigator Handle Paddle
Model:Model: 9730451
Affected:Lot: 402051764
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Brand Name:Medtronic Navigator Percutaneous Reference Pin
Model:Part Numbers: 9733235 & 9733236
Affected:Lots: 2012110744, 2013010347, 2012081458 & 2013041109
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Brand Name:Medtronic Navigator accessories
Model:Part Numbers: 9730950, 9730951, 9731132, 9733497
Affected:Lots: B135276281, B155300281, R683, 121130B respectively
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Brand Name:Medtronic Navigation Thoracic Tactile Probes
Model:Part Number: 9731427
Affected:Lot Numbers: 6688089, 6095124, 6095129
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Brand Name:Medtronic Navigator Sterile Drill Attachments
Model:Part Number: 9733516
Affected:Lot Number: 130204G
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Brand Name:Medtronic Navigator Passive Spheres
Model:Part Numbers: 8801071 & 8801075
Affected:Lot Numbers: 5 lots & 24 lots respectively
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Northern Digital Inc
Brand Name:Medtronic Navigation Bags & Drapes
Model:Part Number: 9732315 & 9732722
Affected:Lot Number: 5246110712, 5316560113 & 6 lots respectively
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Xomed
Brand Name:Medtronic Navigator Biopsy Needle Kit
Model:Part Number: 9733068, 9733065, 9733066
Affected:4 lots, 5 lots, 6 lots respectively
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Issue:A review of the Medtronic internal returned goods procedures showed that there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage. This review identfied the sterile part numbers were processed through the returned goods procedures. A risk analysis was subsequently performed showing that, since there have been no complaints related to sterilisation/package damage there was no change in the likeihood of this scenario leading to harm. Although there have been no reports of issues related to package sterile barrier damage, manufacturer has elected to issue this safety notice to address this possibility.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/5/2014
Hide menus