Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16613 |
| Brand Name: | TCVS Terumo/Sarns Disposable Centrifugal Pump |
| Model: | 164275 |
| Affected: | Lot No:QPO2 |
| Software version: | |
| Recalling Organisation: | Terumo Australia Pty Ltd (NZ) c/- Healthcare Logistics 56 Carrington Road Mount Albert Auckland 1025 |
| Contact Information: | 0800 66 77 57 |
| Manufacturer: | Terumo Cardiovascular Systems Corp |
| Issue: | Manufacturer has received complaints of the Sarns Disposable Centrifugal Punps leaking due to a crack in the molded separator. In all but one report, the leaks were discovered during priming of the cardiopulmonary (CPB) circuit and prior to initiating CPB. In these instances the centrifugal pump was successfully changed out with no delays in the surgical procedures. In one report the failure occurred during CPB and the pump was not replaced. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 1/5/2014 |