Detail of Recall

Type of Product:Device
Medsafe Reference:16613
Brand Name:TCVS Terumo/Sarns Disposable Centrifugal Pump
Model:164275
Affected:Lot No:QPO2
Software version:
Recalling Organisation:Terumo Australia Pty Ltd (NZ)
c/- Healthcare Logistics
56 Carrington Road
Mount Albert
Auckland 1025
Contact Information:0800 66 77 57
Manufacturer:Terumo Cardiovascular Systems Corp
Issue:Manufacturer has received complaints of the Sarns Disposable Centrifugal Punps leaking due to a crack in the molded separator. In all but one report, the leaks were discovered during priming of the cardiopulmonary (CPB) circuit and prior to initiating CPB. In these instances the centrifugal pump was successfully changed out with no delays in the surgical procedures.
In one report the failure occurred during CPB and the pump was not replaced.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/5/2014