Detail of Recall

Type of Product:Device
Medsafe Reference:16612
Brand Name:StimPod NMS 450 Nerve Stimulator
Model:XT-45011, XT-45011-NA, XT-45001 and XT-45001-NA
Affected:
Software version:
Recalling Organisation:Anatech Medical Ltd
P O Box 0171
9 Pirihi Road
Takahiwai
Ruakaka
Whangarei
Contact Information:
Manufacturer:Xavant Technology
Issue:In isolated incidences (less than 0.003%) 0 while using the device in NMBA monitoring mode, electrodes used may deteriorate to a level where current densities may be large enough to cause superficial burns on the skin of the patient. The occurrence of this phenomena is due to a lack of training on the impedance monitoring fucntion of the device and it is imperative that users are properly trained to monitor the impedance of the electrode/skin impedance through the actual current delivered function in MNBA with all Stmpof NMS450s.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/5/2014