Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16612 |
| Brand Name: | StimPod NMS 450 Nerve Stimulator |
| Model: | XT-45011, XT-45011-NA, XT-45001 and XT-45001-NA |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Anatech Medical Ltd P O Box 0171 9 Pirihi Road Takahiwai Ruakaka Whangarei |
| Contact Information: | |
| Manufacturer: | Xavant Technology |
| Issue: | In isolated incidences (less than 0.003%) 0 while using the device in NMBA monitoring mode, electrodes used may deteriorate to a level where current densities may be large enough to cause superficial burns on the skin of the patient. The occurrence of this phenomena is due to a lack of training on the impedance monitoring fucntion of the device and it is imperative that users are properly trained to monitor the impedance of the electrode/skin impedance through the actual current delivered function in MNBA with all Stmpof NMS450s. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/5/2014 |