Type of Product:Device
Medsafe Reference:16604
Brand Name:Philips Brilliance CT series and Ingenuity CT series systems
Model:Brilliance CT 16, 64, Big Bore, iCT, Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Philips has determined there are two potential issues that may result in the subframe of the table top becoming free floating and causing unintended, horizontal motion. The two potential issues are:
1. Service latch may not be properly secured
2. Service latch threaded rod may be fractured
To date, there have been no reported injuries or deaths resulting from these issues.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/5/2014