Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16598 |
| Brand Name: | RaySearch RayStation Radiation therapy treatment planning system 4.0 |
| Model: | |
| Affected: | |
| Software version: | Version 4.0.3.4 |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | There is a potential problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI (region of interest) can be wrong. This affects all dose-volume properties for such ROIs, including DVH (dose-volume histogram), dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/4/2014 |