Detail of Recall

Type of Product:Device
Medsafe Reference:16596
Brand Name:Synthes Craniomaxillofacial (CMF) Distraction System, BC Distraction Bodies
Model:Models: 04.315.064 and 04.315.066
Affected:Lots: 6342004, 7351857 and 6651388, 6913921, 718588 respectivly
Software version:
Recalling Organisation:Synthes New Zealand Ltd
507 Mt Wellington Highway
Mt Wellington
Auckland 1060
Contact Information:0800 657 734
Manufacturer:Synthes GmbH
Issue:The AB and BC Distractor Bodies used in the Craniomaxillofacial Distraction System may reverse post-operatively.
In the presence of inadequate distraction or reversal of a previously achieved distraction distance, the patient may need to undergo prolonged distraction therapy. In some instances, revision surgery may be needed to replace the device.
Reversing occurs when the distractor screw is turned in the opposite direction to cause the assembly to lose distraction distance.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional