Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16594
Brand Name:Acumed Tension Band Pin
Model:TB-1550K-S TB-1570K-S TB-1590K-S
Affected:Manufactured between Jan 2009 and March 2011
Software version:
Recalling Organisation:Ortho Medics Limited
586 Great South Road
Ellerslie
Auckland 1051
Contact Information:09 527 1203
Manufacturer:Acumed LLC
Issue:This implantable product was manufactured wiht a grade of stainless steel that is different than from what was specified. Cytotoxicity tests confirm that product is NOT cytotoxic. Additional biocompatibility testing is currently underway. To date there have been no reported injuries and one malfunction with no harm to patient. The root cause of the malfunction is as yet inknown.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:28/4/2014
Hide menus
Show menus