Detail of Recall

Type of Product:Device
Medsafe Reference:16577
Brand Name:Medical Intelligence HexaPOD evo RT Couchtop Securing Screws
Model:
Affected:
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Medical Intelligence Medizintechnik GmbH
Issue:If the screws that secure the HexaPOD evo RT Couchtop are not adequatey tightended during installation, they may loosen over tme. This condition has the following impacts:
1. There may be incorrect patient respositioning if the HexaPOD evo RT Couchtop is moved.
2. The HexaPOD evo RT Couchtop could tip when a patient is in a highly extended position (which could be the case in treatments of the lower body half)
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/4/2014