Detail of Recall

Type of Product:Device
Medsafe Reference:16575
Brand Name:Biosense ThermoCool SF Nav Bi-Directional Catheter
Model:
Affected:All lots
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Biosense Webster Inc
Brand Name:Biosense ThermoCool SF NAV Uni-directional Catheter
Model:
Affected:All lots
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Biosense Webster Inc
Issue:The manufacturer has revised the procedural steps and precautions set out in the IFU
1. When using the devices with conventional systems (using fluroscopy to determine catheter tip location) or with the Carto EP Navigations System, careful catheter manipulation must be performed in order to avoid cardiac damage, perforation or tamponade.
2. Do not use the temperature sensor to monitor tissue temperature.
3. Precautionary measures should be taken when ablating on the posterior wall of the left atrium in proximity to the oesophagus to minimse risk of atrio-esophageal fistula.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/5/2014