Detail of Recall

Type of Product:Device
Medsafe Reference:16566
Brand Name:Hamilton G5 or Hamilton -S1 Ventilator
Model:
Affected:
Software version:between V2.00 and V2.31
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Hamilton Medical AG
Issue:Investigations have shown that ventilation and aleams of a Hamilton-G5 or Hamilton-S1 ventilator can be suppressed unintentionally after the activation of a suctioning manoeuvre by the operator; this situation can occur regardless of the selcected patient group (neonatal, pediatric and adult). Ventilation must then be manually restarted.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/5/2014