Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16562
Brand Name:Philips MDC PACS Release R2.x and Intellispace PACS DCX R3.x
Model:
Affected:
Software version:all versions
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Healthcare Informatics Inc
Issue:Manufacturer has identified a software defect that results in incorrect display of absolute values when measuing pixel density and statistics for Region of Interest. This issue occurs when Enhanced MR and Enhanced CT images are sent to the PACS with modality LUT (Look Up Tables) as defined in the DICOM functional group.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/4/2014