Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16559 |
| Brand Name: | Siemens SOMATOM Emotion 6/12 Slice Configuration |
| Model: | |
| Affected: | 10045692, 10165566, 10165888, 10165977, 10415118, 10415119, 10415116 and 10415117 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Shanghai Medical Equipment Ltd |
| Brand Name: | Siemens SOMATOM Spirit/ Spirit Power |
| Model: | |
| Affected: | 10045692, 10165566, 10165888, 10165977, 10415118, 10415119, 10415116 and 10415117 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Shanghai Medical Equipment Ltd |
| Issue: | Manufcaturer confirms detection of a rubber component with unfavorable ageing properties. The mechanical function of this rubber damper, which is situated within the mounting assembly of the ritation motor, may potentially degrade over time and may impact stability. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/4/2014 |